930i Vitrek AC Grounding HiTESTER
Alternating Current Output
Range 3 to 30 amperes RMS, programmable
Resolution 1 amp
Output Frequency Sinewave synthesis, 50 Hz to 400Hz
Compliance Voltage 3 volts RMS
Regulation ± 0.25% no load to full load
Resistance Measurement
Resistance Limit User programmable max. and min. Test is automatically aborted when limit is exceeded.
Resistance Range 1Ω at 3 amps linearly decreasing to 100 mΩ at 30A
Resistance Resolution 10mΩ
Measurement Accuracy ± 1.0% of reading; ± 0.5% of range
Accuracy Specifications apply for 1 year, +/- 5°C from the cal temp within 20°C to 30°C
Calibration Certificate of calibration traceable to the National Institute of Standards and Technology (NIST) provided
Test Time Digital count down timer, programmable 1 to 9999 seconds
Terminal Configuration True four-wire kelvin
General Specifications
Remote Interface VICL and serial com standard. PLC optional.
Display 40 character dot matrix illuminated liquid crystal display
Indicators L.E.D. status indicators
Alarm Audible and visual alarms provide pass/fail indication
Diagnostics Built-in self diagnostics test ROM, RAM and output system
Safety CE mark certified to EN61010
Accessories K-2, VIC-3, BBL-2 and operators manual supplied with unit
Warranty Extended 5 year parts and labor with annual factory calibration and maintenance. Standard warranty one year.
Operating Temperature 0°C to 50°C
Humidity 90% RH 0°C to 40°C
Power 115/230 VAC +/- 10% 250 VA max
Dimensions 89mmH x 432 mmW x 457mmD (3.5"H x 17"W x 18"D)
Weight 11Kg (24 lb.) Net/15Kg (34 lb.) shipping
REQUEST A QUOTE