TOP
FSH 휴대용 스펙트럼 분석기
R&S®FSH Handheld Spectrum Analyzer: 현장에서 설치 및 유지보수를 위한 올인원 핸드헬드 플랫폼. 주파수 대역은 9 kHz ~ 3.6/8/13.6/20 GHz입니다. 모델에 따라 내부 Tracking Generator 및 VSWR 브릿지가 함께 제공될 수 있습니다.
REQUEST A QUOTE
Detail
|
Spec
|
Option
|
Download
|
Video
Detail
  • 9 kHz ~ 최대 20 GHz
  • < –161 dBm (1 Hz)의 뛰어난 감도(프리앰프 사용 시)
  • 올인원 분석기: 스펙트럼 분석기, 네트워크 분석기 및 CAT
  • 내부 추적 발생기, VSWR 브릿지, 바이어스 티
  • 생활방수 기능의 견고한 현장용 하우징
Spec
Spectrum analysis
 R&S®FSH4R&S®FSH8R&S®FSH13R&S®FSH20
Frequency rangemodels .04/.14/.08/
.18/.13/.23/.20/.30
9 kHz to 3.6 GHz9 kHz to 8 GHz9 kHz to 13.6 GHz9 kHz to 20 GHz
 models .24/.28100 kHz to 3.6  GHz100 kHz to 8 GHz
Resolution bandwidths 1 Hz to 3 MHz
Displayed average noise levelwithout preamplifier, RBW = 1 Hz (normalized)
 9 kHz to 100 kHz
(models .04/.14/
.08/.18 only)
< –108 dBm, –118 dBm typ.< –96 dBm, –106 dBm typ.
 100 kHz to 1 MHz< –115 dBm,  –125 dBm typ.
 1 MHz to 10 MHz< –136 dBm,  –144 dBm typ.
 10 MHz to 2 GHz< –141 dBm,  –146 dBm typ.
 2 GHz to 3.6 GHz< –138 dBm,  –143 dBm typ.
 3.6 GHz to 5 GHz< –142 dBm, –146 dBm typ.
 5 GHz to 6.5 GHz< –140 dBm, –144 dBm typ.
 6.5 GHz to 13.6 GHz< –136 dBm, –141 dBm typ.
 13.6 GHz to 18 GHz< –134 dBm,  –139 dBm typ.
 18 GHz to 20 GHz< –130 dBm, –135 dBm typ.
 with preamplifier, RBW = 1 Hz (normalized)
 100 kHz to 1 MHz< –133 dBm,  –143 dBm typ.
 1 MHz to 10 MHz< –157 dBm,  –161 dBm typ.< –155 dBm, –160 dBm typ.
 10 MHz to 2 GHz< –161 dBm, –165 dBm typ.
 2 GHz to 3.6 GHz< –159 dBm,  –163 dBm typ.
 3.6 GHz to 5 GHz< –155 dBm,  –159 dBm typ.
 5 GHz to 6.5 GHz< –151 dBm, –155 dBm typ.
 6.5 GHz to 8 GHz< –147 dBm, –150 dBm typ.
 8 GHz to 13.6 GHz< –158 dBm,  –162 dBm typ.
 13.6 GHz to 18 GHz< –155 dBm,  –160 dBm typ.
 18 GHz to 20 GHz< –150 dBm,  –155 dBm typ.
Third-order intercept (IP3)300 MHz to 3.6  GHz> +10 dBm,  +15 dBm typ.
 3.6 GHz to 20 GHz> +3 dBm,  +10 dBm typ.
Phase noisefrequency 500 MHz
 30 kHz carrier offset< –95 dBc (1 Hz), –105 dBc (1 Hz) typ.
 100 kHz carrier offset< –100 dBc (1 Hz),  –110 dBc (1 Hz) typ.
 1 MHz carrier offset< –120 dBc (1 Hz),  –127 dBc (1 Hz) typ.
Detectorssample, max. peak, min. peak, auto peak, RMS
Level measurement uncertainty10 MHz < f ≤ 3.6 GHz<1 dB,  0.5 dB typ.
 3.6 MHz < f ≤ 20 GHz<1.5 dB,  1 dB typ.
Display 6.5" color LCD with VGA resolution
Battery operating time (without tracking generator)R&S®HA-Z204, 4.2 Ahup to 3 h
 R&S®HA-Z206, 6.3 Ahup to 4.5 h
Dimensions (W × H × D) 194 mm × 300 mm × 69 mm (144 mm)1)
7.6 in × 11.8 in × 2.7 in (5.7 in)
Weight 3 kg (6.6 lb)
Vector network analysis2)/vector voltmeter3)
 R&S®FSH4R&S®FSH8R&S®FSH13/20
Frequency rangemodels .24 /.28 /.23 /.30300 kHz to 3.6 GHz300 kHz to 8 GHz100 kHz to 8 GHz
Output power (port 1) 0 dBm to –40 dBm-
Output power (port 2) 0 dBm to –40 dBm0 dBm to –40 dBm
Reflection measurement (S11, S22)
Directivity300 kHz to 3 GHz> 43 dB nominal> 43 dB nominal> 43 dB nominal 4)
 3 GHz to 3.6 GHz> 37 dB nominal> 37 dB nominal> 37 dB nominal 4)
 3.6 GHz to 6 GHz> 37 dB nominal> 37 dB nominal 4)
 6 GHz to 8 GHz> 31 dB nominal> 31 dB nominal 4)
Display modesvector reflection and transmission measurement (R&S®FSH-K42)magnitude, phase, magnitude + phase, Smith chart, VSWR, reflection coefficient, mρ, on-port cable loss, electrical length, group delay
 vector voltmeter (R&S®FSH-K45)magnitude + phase, Smith chart
Transmission measurements
Dynamic range (S21)100 kHz to 300 kHz70 dB typ.70 dB typ.-
 300 kHz to 3.6 GHz> 70 dB, 90 dB  typ.> 70 dB, 90 dB  typ.-
 3.6 GHz to 6 GHz> 70 dB, 90 dB  typ.-
 6 GHz to 8 GHz50 dB typ.-
Dynamic range (S12)100[nbsp]]kHz to 300 kHz80 dB typ.80 dB typ.typ. 80 dB
 300 kHz to 3.6 GHz> 80 dB, 100 dB typ.> 80 dB, 100 dB typ.80 dB, 100 dB typ.
 3.6> 80 dB, 100 dB typ.80 dB, 100 dB typ.
 6 GHz to 8 GHz60 dB typ.60 dB typ.
Display modesvector reflection and transmission measurement (R&S®FSH-K42)magnitude (attenuation, gain), phase, magnitude + phase, electrical length, group delay
 vector voltmeter (R&S®FSH-K45)magnitude + phase
Option


Download
Video